4.5.4 Hold function: “freezing” of outputsfreezes” the outputs
The current output is “frozen” in the setup mode and during calibration, i.e. the last current
value is constantly output. The display shows the “HOLD” message. In case of steady control
(4... 20 mA) on current output 2, it is set to 0/4 mA during Hold.
Note:
• Hold settings can be found in
chapter 5.5, function S2.
• During automatic operation, all
contacts will go to their normal
positions.
• An active hold has priority over all
other automatic functions.
• With every hold, the I component
of the controller is set to zero.
• A possibly accumulated alarm
delay is reset to “0".
• The hold function can also be
activated externally via the hold
input (see wiring diagram in
Fig. 3.10; digital input 1).
• The manual hold (field S3) remains
active even after a power failure.