Coding Field
Selection or range
Factory setting (bold)
Display Info
F6
Set alarm contact
to be effective for
selected error
yes
no
If set “no” all the other alarm
settings (e. g. alarm delay) are
also deactivated. The settings
themselves are retained.
This setting only applies to the
error selected in F5.
Factory setting is no starting with
E080.
F7
Set error current
to be effective for
selected error
no
yes
The error selected in F4
becomes effective or is
suppressed when it occurs.
This setting only applies to the
faults selected in F5.
F8
Automatic start of
cleaning function?
no
yes
This field only exists for some
errors, see chapter 7.1.
F9
Return to menu or
select next error
Next = Next error
<—-R
If <—-R is selected, the display
returns to F; if next is selected
the display returns to F5.
Factory settings are printed in bold face;
base version does not include functions in
italic
.
yes F6
Rel.Assg
no F7
Curr.Assg
no F8
CleanTrig
F9
Select
next
Instrument configuration Liquisys M CPM 223 / 253
34 Endress+Hauser