5.4 Current outputs
The function group CURRENT OUTPUT is used to configure the individual outputs. Either a
linear (O3 (1)) or, in conjunction with the plus package, a user-defined current output
characteristic (O3 (3)) can be entered.
Furthermore, a current output value can be simulated to check the current outputs
(O3 ( 2)).
The controller set value in field R 237 / R 266 can be output via current output 2, if available.
The distance ∆ signal per mA between two table value pairs must exceed:
• pH: 0.03
• Redox: 5 mV
• Temperature: 0.25 °C
First enter the current output configuration you require in the following blank table. Ensure the
required minimum distance by calculating the resulting signal distance per mA. Then enter the
result in the instrument.
10
8
12
6
0 48 121620
pH
Current [mA
∆ current
Fig. 5.7
User-defined current output
characteristic
Current output 1 Current output 2
Value pair
pH / mV /
%/°C
[]
Current
[mA]
Distance
per mA
pH / mV /
%/°C
[]
Current
[mA]
Distance
per mA
1
2
3
4
5
6
7
8
9
10
Instrument configuration Liquisys M CPM 223 / 253